1 | Bomb capacity | (300 ± 50) mL |
2 | Oxygen bomb pressure range | 0.5 ~ 3.5MPa, no leakage. |
3 | Crucible | Φ 25mm, height 14-19 mm,thickness 1 mm. Stainless steel material |
4 | Timer | accuracy of 0.5s |
5 | Maximum power | 0.5KW |
6 | Ambient temperature | (10 ~ 35) ºC, humidity ≤80%; |
7 | Heat capacity of the instrument | about 10000J / K |
8 | Heat repeatability error | ≤0.2% |
9 | Temperature range | (4.5-50) ºC |
10 | Temperature detection | 0.0001 ºC resolution, using the PT100 sensor |
11 | Inner cylinder volume | 2000ml; Oxygen bomb Sealing: oxygenation pressure 2.5 ~ 3.5MPa, no leakage |
12 | Stirring power | 3W; Strring speed: 375r/min |
13 | Ignition time | 5s |
14 | Ignition voltage | DC 12V |