After-sales Service: | Response in 12 Hours |
---|---|
Warranty: | 12 Months |
Power Supply: | AC 220V±10% |
Customization: |
---|
Shipping Cost:
Estimated freight per unit. |
about shipping cost and estimated delivery time. |
---|
Payment Method: | |
---|---|
Initial Payment Full Payment |
Currency: | US$ |
---|
Return&refunds: | You can apply for a refund up to 30 days after receipt of the products. |
---|
Suppliers with verified business licenses
Introduction
The test method of Cavitation Corrosion and Erosion-Corrosion Characteristics Tester is: pump the sample with a standard aluminum automobile water pump, run it for 100h under the conditions of 113 ºC, 103kpa and pump speed of 4600r / min, and then check the water pump to determine the degree of cavitation corrosion, and grade it according to the grading system table.
Characteristic
1. The Cavitation Corrosion and Erosion-Corrosion Characteristics Tester is composed of operation and display, heating control, speed control, pressure control, flow control system and supporting components;
2. Motor: used for coolant circulation drive, linked with water pump;
3. Water pump: the standard ASTM D method requires imported water pump;
4. Vacuum pressure gauge: digital display of real-time pump inlet vacuum;
5. Throttle valve: used to regulate the flow rate of water pump;
6. Heater: heat the coolant to the test temperature;
7. Liquid storage tank: stores cooling liquid for pipeline system;
8. Heat exchanger: provide 35 ~ 38 ºC test temperature through cooling water heat exchange;
9. Regulating valve: three regulating valves are used to adjust vacuum degree and test pressure;
10. Pressure gauge: displays the pressure during the test;
11. Expansion tank: steam expansion device
Technical specifications of ASTM D2809 Cavitation Corrosion and Erosion-Corrosion Characteristics Tester
1 | Applicable standards | ASTM D2809, SH/T0087 |
2 | Environmental requirements | 20 ºC ~ 25 ºC, humidity < 60% |
3 | Fuse | 20A |
4 | Pump speed | 0 ~ 4600 ± 100r / min |
5 | Temperature control range | < 650 ºC |
6 | Resolution (measurement) | 0.1 ºC |
7 | Temperature control accuracy | ± 0.1 ºC |
8 | Flow range | 0 ~ 20ml / min |
9 | Flow control accuracy | ± 5% |
10 | Power supply | AC220V 50Hz |
11 | Power consumption | 6.5KW |
FAQ