Introduction:
The flammability chamber tester is designed according to Federal Aviation Regulation Sec. 25.853 and CCAR-25 Appendix F of the first part b (4), b (6), b (7) .The flammability test chamber is integrated appendix F the first part b4 vertical combustion and b 6 45 ° and 60 ° b7 item combustion demand for housing and design ;And b2 is the request of the sample holder and customized for various experiment using the sample holder and corresponding independent test procedure.
The flammability chamber tester is designed with touch screen control system and PLC, it can make the experiment more automaticly and easier to extract data; and provide the combustion at the angle of 90°, 45 ° and 60 ° with independent test procedures;
Standard:
· FAR.25.853:
Department of transportation, Federal Aviation administration regulations for compartment interiors
· CCAR-25 Appendix F of the first part of b 4 and 6,7 items
Main features:
1.The tank is painted and equipped with viewing window;
2. The chamber dimensions: 800mm x 450mm x 640mm;
3. The Flammability Chamber Tester has function of automatic electronic ignition;
4. Ignition timing can be arbitrarily set around 0 ~ 99.99 seconds, continued burning time and flame time can be automatically recorded and digitally displayed;
5, The flame height can be adjusted;
6. The "push-pull" fixture is supplied with Flammability Chamber Tester;
7. The burner is located at the sliding track, easily operated, when pushed to the combustion position, automatic timing.
8. Needle valve can precisely control the gas flow and adjust standard flame height around 38.1mm;
9. According to the test samples, multiple samples holder can be chosen; customizing BWF sample size and sample holder.
10. Brand pressure gauge and pressure regulator can accurately adjust the gas pressure;
11. Bunsen burner tube inner diameter is 9.5mm; the bottom of sample must be higher than the light of 19.1mm
12.Thermocouple can test the maximum flame temperature about 1000 ºC; flame kerne temperature must be higher than 843 ºC.
13.The ignition time can be set freely and the buzzer can remind people when the time is over;
14. PLC microcomputer control system with touch screen can make the experimental process more automaticly and the data can be automatically recorded and stored;
Specification
1 |
The ignition timing |
0 to 99.99 seconds set arbitrarily |
2 |
After flame time and anti flaming time recorder |
0 ~ 99.99 seconds, accuracy ± 0.01 seconds |
3 |
The flame height measuring device |
flame height ruler is about 38.1mm |
4 |
Gas sources |
methane, propane,butane, liquefied petroleum gas or other flammable gases |
5 |
Power supply |
AC220V, 50Hz, 50W |
6 |
Weight |
70KG |
Factory overview
Customer Feedback
Germany
USA
Transporation
FAQ
1.What's the delivery time?
About 30 working days.
2.What's your payment terms?
30% paid by T/T in advance,70% paid by by T/T before shipping.
3.What's your guarantee terms?
The warranty of all products is 1 year.During this period,we are responsible for all faults caused by us.And we can supply life-long technical support.
4.Can you supply training?
We can supply training in our factory and online training for free.If onsite installation and training are required,please pay for tickets,accommodation and salary USD150/Day for every person.