Customization: | Available |
---|---|
After-sales Service: | Response in 12 Hours |
Warranty: | 12 Months |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Summary
This Oxidize Resistance Test Instrument is suitable for evaluating the antioxidant characteristics of brake fluid.
The principle is to mix used benzoyl peroxide, distilled water, and brake fluid into a test sample, and place it with 1/8 rubber bowl and aluminum and cast iron pieces. After keeping it in the instrument for 168 hours, take it out and check for any pitting, roughness, and other corrosive phenomena on the test pieces, and calculate the mass change of the test pieces.
Main Features
1. The glass test tube meets the standard requirements, with a diameter of 22mm and a height of 175mm.
2. The heating chamber has ventilation holes on the top.
3. Comes with cast iron pieces, 80mm long, 13mm wide, and 2.5mm thick, and aluminum pieces, 80mm long, 13mm wide, and 2.5mm thick.
4. Comes with tin foil, with a thickness of 0.02~0.06mm, and a tin content of 99.9%.
5. The microcomputer temperature controller has PID self-tuning, PT100 temperature sensor, and high precision.
6. The digital timer controller has a buzzer alarm to remind when the test is finished.
7. The instrument casing is made of cold-rolled steel plate, and the surface is treated with electrostatic spraying.
Technical specifications
1 | Power supply | AC 220 V±10%, 50 Hz |
2 | Temperature control range | 20ºC~150ºC |
3 | Heating method | Electric heating tube heating |
4 | Cooling method | Compressor |
5 | Temperature control method | Digial PID temperature controller |
6 | Timing method | Digital timer |
7 | Alarming method | Buzzer |
8 | Cast iron and aluminum pieces | 80mm in length, 13mm in width, and 2.5mm in thickness |
9 | Glass test tube | φ22mm*175mm |