Introduction:
The Cone Calorimeter offers a method for assessing the heat release rate and dynamic smoke production rate of specimens exposed in the horizontal orientation to controlled levels of irradiance with an external igniter.
The heat release rate is determined by the measurement of the oxygen consumption derived from the oxygen concentration and the flow rate in the combustion gas.
The dynamic smoke production rate is calculated from the measurement of the attenuation of a laser light beam by the combustion gas.
The heat release rate and smoke obscuration are recorded for the entire test, regardless of whether the specimen is flaming or not.
The following date will be reported by the cone calorimeter test:
• Heat release rate, in MJ/kg, kW/m2
• Total heat release, in MJ
• Smoke production rate, in m2/m2
• Total smoke production, in m2
• Mass loss rate, in g/m2.s
• Time to sustained flaming and extinguished, in seconds
• Heat flux, in kW/m2
• Exhaust duct flow rate, in l/s
• Test duration, in seconds
• C factor, in 1/kg.m.
• K Results derived from different standards.
Standards:
• ISO 5660: Reaction to fire tests - the rate of heat release, smoke generation rate and mass loss rate.
• ASTM E 1354: Standard Test Method of materials heat and visible smoke release rate.
• BS 476 Pt.15: Building materials and structures test: - Test method for heat release rate of products.
• ASTM E1474:Standard Test Method for Determining the Heat Release Rate of Upholstered Furniture and Mattress Components or Composites Using a Bench Scale Oxygen Consumption Calorimeter.
• ASTM E1740:Standard Test Method for Determining the Heat Release Rate and Other Fire-Test-Response Characteristics of Wall Covering or Ceiling Covering Composites Using a Cone Calorimeter.
• ASTM F1550:Standard Test Method for Determination of Fire-Test-Response Characteristics of Components or Composites of Mattresses or Furniture for Use in Correctional Facilities after Exposure to Vandalism, by Employing a Bench Scale Oxygen Consumption Calorimeter.
• ASTM D6113:Standard Test Method for Using Cone Calorimeter to Determine Fire-Test-Response Characteristics of Insulating Materials Contained in Electrical or Optical Fiber Cables.
• CAN ULC 135:STANDARD TEST METHOD FOR THE DETERMINATION OF COMBUSTIBILITY PARAMETERS OF BUILDING MATERIALS USING AN OXYGEN CONSUMPTION CALORIMETER (CONE CALORIMETER).
Main Features
1.Cone-shaped radiant electrical heater, an electrical heater rod tightly wound into a truncated cone, rated 5000w with a producing irradiance output of 100 kW per square meter.
2.Horizontally orientated specimen mounting facility.
3.PID temperature control of cone irradiance.
4.Load cell, having a resolution of 0.1g and accuracy better than ±0.3g.
5.Well heat insulation under the heating cone.
6.Independent marble cone facility workbench.
7.Exhaust gas system, stainless steel made for long life use, adjustable flow control from 0-50g/s at a resolution of 0.1g/s.
8.An orifice plate flow meter to measure the combustion gas flow in the duct.
9.Combustion gas sampling apparatus, consisting of 3-stages filters (accuracy up to 0.5 microns), refrigerated cold trap, diaphragm pump, moisture trap.
10.Oxygen analyzer, paramagnetic type with the range of 0 to 25%.
11.SIEMENS® or ABB® gas analyzer is optional.
12.Carbon monoxide and carbon dioxide analyzer, NDIR type with the range of 0-1% & 0-10%.
13.SIEMENS® or ABB® gas analyzer is optional.
14.Smoke obscuration measuring system, consisting of He-Ne laser generator, silicon photodiodes, and mounting facility.
15.Methane mass flow controller for calibration, accuracy up to ±1.0% F.S.
16.Heat flux meter, with the self-recycle cooling system.
17.Data acquisition system, maximum record the data 50 times per second.
18.Operation software, with the functions of the standard test program, sensor calibration, auto system calibration, status monitoring, and report management.
Technical Specifications:
1.Power requirement: 220 V, 50 Amps
2.Environmental temperature: 10 ° C to 35 ° C
3.Working gas: Air compressor gas
4.Calibration gases: methane, nitrogen, CO2/CO mixture gas
5.Size: 2250 * 1000 * 2230MM3
6.Weight: 538kg