Customization: | Available |
---|---|
After-sales Service: | Response in 12 Hours |
Warranty: | 1 Year |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Introduction
The Non-combustibility Test Instrument is used for determining the non-combustibility performance of the building and railway material under a 750ºC furnace.
During the Non-Combustibility Test, the test specimen (diameter 45mm and height 50mm) is inserted into a stabilized furnace at 750ºC, and the temperature of the furnace, surface & center of the specimen will be measured. Also, the flaming time and moss loss are recorded.
Non-Combustibility Test Instrumentis widely used to assess the flammability performance class of materials, such as EN 13501-1, EN 45545-2, DIN 4102, IMO FTP Code, etc.
1.Special furnace support with Teflon finish for anti-corrosion.
2.A refractory tube furnace, the operating temperature resisting over 1000 degrees.
3.The Non-combustibility Test Instrument is equipped with Kanthal® heating element, provides a steady performance.
4.The heating element is replaceable and easy to maintain.
5.The furnace conforms to various international standards:
(Two furnace thermocouples-ISO1182:2020,or one furnace thermocouple).
6.The The Non-combustibility Test Instrument is equipped with Omega® thermocouples, and provide spare parts.
7.Automated furnace temperature control system, provides a fast temperature stabilization program and calibration stabilization program.
8.CE-certificated PID temperature controller.
9. Standard test program and Individual furnace temperature calibration program.
10.Stainless steel made sample holder.
11.Fast specimen holder loading fixturing.
12.Data acquisition system, record frequency once per second.
13.Temperature solution at 0.1ºC, timing solution at 0.1sec.
14.Report management and export
Electrical |
220V, 50Hz, 10A |
Ambient Temperature | 10°C to 35°C |
Dimensions | 1180*750*1350MM3 |
Weight | 65KG |
Applicable standards |
ISO 1182, BS 476-4 & 11,ASTM E 136,IMO FTPC Part 1; ASTM E2652 |