IEC61241 Minimum Ignition Temperature Tester for Electrical Products
| Customization: | Available |
|---|---|
| After-sales Service: | Response in 12 Hours |
| Warranty: | 12 Months |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Find Similar Products
Basic Info.
- Model NO.
- AIC-61241
- Test Standard
- IEC61241,En 50281-2-1
- Size
- 500X300X450 mm
- Keyword
- The Minimum Ignition Temperature Tester
- Test Temperature Range
- Room Temperature ~ 800 ℃
- Display
- LCD
- Power Supply
- 230V / 50Hz or 110V / 60Hz.
- Furnace Working Temperature Range
- Room Temperature ~ 800 ℃
- Temperature Accuracy
- ±1% (Above 500℃)
- Disperse Air Chamber Pressure
- 0 ~ 1 Bar
- Color
- Optional
- Pressure Accuracy of Gas Storage Device
- ±5%
- Warrant
- 1 Year
- Transport Package
- Plywood Case
- Specification
- Stainless steel
- Trademark
- AI
- Origin
- China
- HS Code
- 9031809090
- Production Capacity
- 5 Sets/Month
Product Description
Introduction
The Minimum Ignition Temperature Tester is a way of testing the sensitivity of the ignition that the dust cloud is in the heated environment. When a lot of dust is exposed in heating the air and the air temperature is high enough, it may lead to spontaneous combustion. This method can test the minimum ignition temperature of combustible dust cloud.
Standard:
- IEC 61241-2-1:1994:Electrical apparatus for use in the presence of combustible dust - Part 2: Test methods - Section 1: Methods for determining the minimum ignition temperatures ofdust
- EN 50281-2-1:1999:Electrical apparatus for use in the presence of combustible dust - Part 2-1: Test methods - Methods for determining the minimum ignition temperatures of dust
Technical parameters:
| 1 | Furnace operating temperature range |
Room temperature ~ 800 ºC. The operating temperature can rise to 1000 ºC to clean the test chamber |
| 2 | emperature accuracy | ± 1% (higher than 500 ºC), ± 3% (lower than 300 ºC). |
| 3 | The pressure of the dispersion air chamber | 0 ~ 1 bar |
| 4 | The pressure accuracy of the air storage device | ± 5% |
| 5 | Size | 500X300X450 mm |
| 6 | Power supply | 230v / 50Hz or 110v / 60Hz |
Testing process
The Test is carried on the in Godbert-Greenwald furnace. The heating furnace is consist of the vertical quartz tubes . The hole is at the bottom of the tube.The glass observing room is connected with sample fixing device at the top of the quartz tubes. The compressed air from the tank blows the dust into the heating furnace and the sample fixing device ,and then exposed into the glass observing room and the quartz tubes .the heating furnace is located on the holder . And you can observe the ignition appearance from the bottom of the furnace. The test begins from the 500ºC. If you observe the ignition appearance ,you shall decrease the temperature at the 10ºC,until the ignition does not appear.
Results description
When the ignition occurs , the lowest temperature is recorded as the minimum ignition temperature of the dust cloud .
The minimum ignition temperature of the test data information is used to determine whether the surface temperature of the equipment can cause the automatic fire of the exposed dust . It is an important standard to evaluate the selection of the equipment at the condition of the dusty environment . When using the minimum ignition temperature testing data ,you shall leave a safety limit for the small-scale test .
Factory overview
Customer Feedback
Germany
USA
Transporation
FAQ
About 6 months.
2.What's your payment terms?
30% paid by T/T in advance,70% paid by by T/T before shipping.
3.What's your guarantee terms?
The warranty of all products is 1 year.During this period,we are responsible for all faults caused by us.And we can supply life-long technical support.
4.Can you supply training?
We can supply training in our factory and online training for free.If onsite installation and training are required,please pay for tickets,accommodation and salary USD150/Day for every person.
Send your message to this supplier
People who viewed this also viewed
Find Similar Products By Category
- Supplier Homepage
- Products
- Fire Testing Equipment
- Rail Transit